ICU Medical Recalls Certain Portex Endotracheal Tubes; FDA Issues Early Warning

The FDA Center for Devices and Radiological Health (CDRH) has issued an early warning, and ICU Medical has sent letters to customers recommending discontinuation and removal of certain Portex endotracheal tubes. Affected products include models 100/127/025, 100/127/030, 100/127/035, 100/134/030, and 100/134/035, covering uncuffed Murphy eye oral/nasal tubes and Polar Clear South oral tubes. ICU Medical found that some lots used resin with fluctuating plasticizer ratios, which may make the tubing softer than expected; when used in neonates and early infants, increased ventilation resistance, tube kinking, and difficult intubation may delay intubation, and if not detected during intubation, may lead to inadequate ventilation and unplanned tube changes, and may cause serious adverse events including death. As of November 16, 2026, ICU Medical has reported 1 death associated with this issue.